dc.identifier.citationAkbas U., Kesen N. D. , Koksal C., Bilge H. B. , "Surface and Buildup Region Dose Measurements with Markus Parallel-Plate Ionization Chamber, GafChromic EBT3 Film, and MOSFET Detector for High-Energy Photon Beams", ADVANCES IN HIGH ENERGY PHYSICS, cilt.2016, 2016
dc.identifier.issn1687-7357
dc.identifier.otherav_10875f1e-f35f-4cfa-ba88-7e0b5ad4971d
dc.identifier.othervv_1032021
dc.identifier.urihttp://hdl.handle.net/20.500.12627/16606
dc.identifier.urihttps://doi.org/10.1155/2016/8361028
dc.description.abstractThe aim of the study was to investigate surface and buildup region doses for 6MV and 15MV photon beams using a Markus parallel-plate ionization chamber, GafChromic EBT3 film, and MOSFET detector for different field sizes and beam angles. The measurements were made in a water equivalent solid phantom at the surface and in the buildup region of the 6MV and 15MV photon beams at 100 cm source-detector distance for 5 x 5, 10 x 10, and 20 x 20 cm(2) field sizes and 0 degrees, 30 degrees, 60 degrees, and 80 degrees beam angles. Thesurface doses using 6 MVphoton beams for 10 x 10 cm(2) field sizewere found to be 20.3%, 18.8%, and 25.5% for Markus chamber, EBT3 film, and MOSFET detector, respectively. The surface doses using 15MV photon beams for 10 x 10 cm(2) field size were found to be 14.9%, 13.4%, and 16.4% for Markus chamber, EBT3 film, and MOSFET detector, respectively. The surface dose increased with field size for all dosimeters. As the angle of the incident radiation beam became more oblique, the surface dose increased. The effective measurement depths of dosimeters vary; thus, the results of the measurements could be different. This issue can lead to mistakes at surface and buildup dosimetry and must be taken into account.
dc.language.isoeng
dc.subjectTemel Bilimler (SCI)
dc.subjectFİZİK, PARTİKLER VE ALANLAR
dc.subjectFizik
dc.subjectTemel Parçacıklar ve Alanlar
dc.subjectTemel Bilimler
dc.titleSurface and Buildup Region Dose Measurements with Markus Parallel-Plate Ionization Chamber, GafChromic EBT3 Film, and MOSFET Detector for High-Energy Photon Beams
dc.typeMakale
dc.relation.journalADVANCES IN HIGH ENERGY PHYSICS
dc.contributor.departmentİstanbul Üniversitesi , ,
dc.identifier.volume2016
dc.contributor.firstauthorID26466


Files in this item

FilesSizeFormatView

There are no files associated with this item.

This item appears in the following Collection(s)

Show simple item record