dc.contributor.author | DEMİR, MUSTAFA | |
dc.contributor.author | Isikci, Nazenin Ipek | |
dc.contributor.author | Abuqbeitah, Mohammad | |
dc.date.accessioned | 2022-07-04T16:09:12Z | |
dc.date.available | 2022-07-04T16:09:12Z | |
dc.date.issued | 2022 | |
dc.identifier.citation | Isikci N. I. , Abuqbeitah M., DEMİR M., "The Interference of Gamma Rays With Bone Mineral Density Measurements in Lu-177-PSMA and DOTATATE Therapy", JOURNAL OF CLINICAL DENSITOMETRY, cilt.25, sa.2, ss.237-243, 2022 | |
dc.identifier.issn | 1094-6950 | |
dc.identifier.other | av_d60264e9-6e11-4c64-abb3-b2d951c7f4fb | |
dc.identifier.other | vv_1032021 | |
dc.identifier.uri | http://hdl.handle.net/20.500.12627/184872 | |
dc.identifier.uri | https://doi.org/10.1016/j.jocd.2021.08.003 | |
dc.description.abstract | The main purpose was to describe the interference of gamma radiation emitted by Lu-177 with simultaneous bone mineral density BMD measures for patients undergoing Lu-177-PSMA and Lu-177 DOTATATE therapy. A cohort of 9 patients underwent Lu-177-PSMA therapy were randomly selected to speculate the activity in the abdominopelvic region. So that, SPECT/CT scan at 24 h was used with attenuation and scatter correction. The activities were derived from the delineated ROIs over the abdominopelvic zone showing a range of 34-274 MBq. Next, a water path was placed under spine phantom mimicking L1-L4 vertebrae and followed by consecutive DEXA scans made by Hologic 4500 W and GE-Lunar DPX-NT systems. Five scans were performed without/and with different Lu-177 activities 37, 185, 370 and 555 MBq under the same geometric conditions. The obtained BMD readings of L1-L4 by the Hologic device were 1.027, 1.024, 1.021, 1.013, and 1.006 g/cm(2) with presence of 0, 37, 185, 370, and 555 MBq Lu-177 activity, respectively. Whereas, in Lunar device, it was found as higher as 1.163, 1.121, 1.09, 1.072, and 1.043, respectively. There was no statistically significant difference between both devices (p(value) >= 0.05). The fluctuation ranges in the L1-L4 BMD readings at the presence of 37-555 MBq were 0.3%-2%, and 3.6%-10.3% for Hologic and Lunar systems, respectively. It was emphasized that gamma radiation emitted by Lu-177 relatively influence DEXA scans and the yielded BMD measures. Postponing DEXA scans as early as 8 d after Lu-177-PMSA and 11 d after Lu-177-DOTATATE therapies is recommended to avoid the erroneous contribution of gamma radiation and provide precise bone assessment. | |
dc.language.iso | eng | |
dc.subject | Tıp | |
dc.subject | ENDOKRİNOLOJİ VE METABOLİZMA | |
dc.subject | Klinik Tıp | |
dc.subject | Klinik Tıp (MED) | |
dc.subject | Sağlık Bilimleri | |
dc.subject | Dahili Tıp Bilimleri | |
dc.subject | İç Hastalıkları | |
dc.subject | Endokrinoloji ve Metabolizma Hastalıkları | |
dc.subject | Endocrinology | |
dc.subject | Endocrine and Autonomic Systems | |
dc.subject | Endocrinology, Diabetes and Metabolism | |
dc.subject | Life Sciences | |
dc.subject | Health Sciences | |
dc.title | The Interference of Gamma Rays With Bone Mineral Density Measurements in Lu-177-PSMA and DOTATATE Therapy | |
dc.type | Makale | |
dc.relation.journal | JOURNAL OF CLINICAL DENSITOMETRY | |
dc.contributor.department | Nişantaşı Üniversitesi , , | |
dc.identifier.volume | 25 | |
dc.identifier.issue | 2 | |
dc.identifier.startpage | 237 | |
dc.identifier.endpage | 243 | |
dc.contributor.firstauthorID | 3416952 | |