Basit öğe kaydını göster

dc.contributor.authorGenc, Sema
dc.contributor.authorKucukgergin, Canan
dc.contributor.authorAdemoglu, Evin
dc.contributor.authorOmer, Beyhan
dc.date.accessioned2021-03-05T12:06:43Z
dc.date.available2021-03-05T12:06:43Z
dc.date.issued2019
dc.identifier.citationKucukgergin C., Ademoglu E., Omer B., Genc S., "Performance of automated urine analyzers using flow cytometric and digital image-based technology in routine urinalysis", SCANDINAVIAN JOURNAL OF CLINICAL & LABORATORY INVESTIGATION, cilt.79, ss.468-474, 2019
dc.identifier.issn0036-5513
dc.identifier.otherav_ab7f036e-8401-494e-b151-1fe1e514ab80
dc.identifier.othervv_1032021
dc.identifier.urihttp://hdl.handle.net/20.500.12627/114495
dc.identifier.urihttps://doi.org/10.1080/00365513.2019.1658894
dc.description.abstractThe purpose of this study was to evaluate the analytical performances of Sysmex UF-5000 and Dirui FUS-200 and to compare the results with manual microscopy and between each other. Two hundred fifty urine samples were analyzed for evaluation. Mid-stream specimens were studied sequentially using Dirui FUS-200 and Sysmex UF-5000, and also with manual microscopy within one hour. The physical and chemical components of urinalysis, and sediment results were investigated. The precision results of the FUS-200 and UF-5000 for WBCs, RBCs, and ECs were acceptable. The both analyzers demonstrated good linearity (r > 0.97), with no carry-over. The comparisons of FUS-200 and UF-5000 with manual microscopy for RBCs, WBCs, and ECs on 250 samples exhibited good agreement with little bias (R > 0.780). Only, the moderate agreements were obtained for calcium oxalate for both analyzers (R = 0.512, and 0.648, respectively). The sensitivities of the FUS-200 and UF-5000 were 75.8% and 86.8%, with specificities of 92.3% and 87.8% for WBCs, for RBCs the sensitivities were 91.1%, and 84.4% with specificities of 82.2%, and 89.6% for both analyzers. Kappa values of the UF-5000 were higher than FUS-200 for WBCs, RBCs, ECs, and calcium oxalate. The FUS-200 and UF-5000 urine analyzers, are both accurate, very precise systems and can be safely used in clinical laboratories. However, due to the technological characteristics of the UF-5000 analyzer, its positive impacts on the morphologic recognition and enumeration of RBCs and WBCs should be taken into account, particularly in university hospital laboratories with high patient volumes.
dc.language.isoeng
dc.subjectTıbbi Ekoloji ve Hidroklimatoloji
dc.subjectDahili Tıp Bilimleri
dc.subjectSağlık Bilimleri
dc.subjectTıp
dc.subjectKlinik Tıp (MED)
dc.subjectKlinik Tıp
dc.subjectTIP, ARAŞTIRMA VE DENEYSEL
dc.titlePerformance of automated urine analyzers using flow cytometric and digital image-based technology in routine urinalysis
dc.typeMakale
dc.relation.journalSCANDINAVIAN JOURNAL OF CLINICAL & LABORATORY INVESTIGATION
dc.contributor.departmentİstanbul Üniversitesi , ,
dc.identifier.volume79
dc.identifier.issue7
dc.identifier.startpage468
dc.identifier.endpage474
dc.contributor.firstauthorID54749


Bu öğenin dosyaları:

DosyalarBoyutBiçimGöster

Bu öğe ile ilişkili dosya yok.

Bu öğe aşağıdaki koleksiyon(lar)da görünmektedir.

Basit öğe kaydını göster