Basit öğe kaydını göster

dc.contributor.authorERTAS, M
dc.contributor.authorYAZICI, JALE
dc.contributor.authorOGE, Ali Emre
dc.contributor.authorBaslo, Mehmet Barış
dc.contributor.authorYILDIZ, N
dc.date.accessioned2021-03-05T14:03:57Z
dc.date.available2021-03-05T14:03:57Z
dc.date.issued2000
dc.identifier.citationERTAS M., Baslo M. B. , YILDIZ N., YAZICI J., OGE A. E. , "Concentric needle electrode for neuromuscular jitter analysis", MUSCLE & NERVE, cilt.23, sa.5, ss.715-719, 2000
dc.identifier.issn0148-639X
dc.identifier.othervv_1032021
dc.identifier.otherav_b538ea4f-220e-4b1f-a264-0ac1a37e6f7c
dc.identifier.urihttp://hdl.handle.net/20.500.12627/120688
dc.description.abstractWe used a concentric needle electrode (CNE) with 2 kHz low-cut filter and a single fiber electrode (SFE) in the same subjects for neuromuscular jitter measurement in the extensor digitorum communis (EDC) and orbicularis oculi (OOc) muscles. At the same session, 20 jitter values were obtained from each subject with each electrode. For EDC (during voluntary contraction), mean jitter values with SFE and CNE were 23.4 +/- 8 mu s and 23.3 +/- 8 us in 10 normals; and 56.8 +/- 28 us and 57.4 +/- 33 mu s in 10 myasthenics. For OOc (during electrical stimulation), mean jitter values with SFE and CNE were 17.9 +/- 5 mu s and 16.3 +/- 4 mu s in 11 normal subjects, and 41.2 +/- 29 mu s and 36.7 +/- 27 mu s in 10 myasthenics. For both muscles, the numbers of individual abnormal jitter values with SFE and CNE were highly comparable. Both needles labeled the same patients as having "normal" or "abnormal" neuromuscular transmission. CNE may be an alternative to SFE in neuromuscular jitter analysis. (C) 2000 John Wiley & Sons, Inc.
dc.language.isoeng
dc.subjectSağlık Bilimleri
dc.subjectDahili Tıp Bilimleri
dc.subjectNöroloji
dc.subjectYaşam Bilimleri
dc.subjectTemel Bilimler
dc.subjectYaşam Bilimleri (LIFE)
dc.subjectTıp
dc.subjectSinirbilim ve Davranış
dc.subjectNEUROSCIENCES
dc.subjectKlinik Tıp (MED)
dc.subjectKlinik Tıp
dc.subjectKLİNİK NEUROLOJİ
dc.titleConcentric needle electrode for neuromuscular jitter analysis
dc.typeMakale
dc.relation.journalMUSCLE & NERVE
dc.contributor.department, ,
dc.identifier.volume23
dc.identifier.issue5
dc.identifier.startpage715
dc.identifier.endpage719
dc.contributor.firstauthorID65381


Bu öğenin dosyaları:

DosyalarBoyutBiçimGöster

Bu öğe ile ilişkili dosya yok.

Bu öğe aşağıdaki koleksiyon(lar)da görünmektedir.

Basit öğe kaydını göster