Comparison of simulated and measured spectra from an X-ray tube for the energies between 20 and 35 keV
Yazar
Yucel, E. Barlas
Emirhan, E.
Yücel, Meriç
Bayrak, Arzu
Özben, Cenap Şahabettin
Üst veri
Tüm öğe kaydını gösterÖzet
Design and production of a simple and low cost X-ray imaging system that can be used for light industrial applications was targeted in the Nuclear Physics Laboratory of Istanbul Technical University. In this study, production, transmission and detection of X-rays were simulated for the proposed imaging device. OX/70-P dental tube was used and X-ray spectra simulated by Geant4 were validated by comparison with X-ray spectra measured between 20 and 35 keV. Relative detection efficiency of the detector was also determined to confirm the physics processes used in the simulations. Various time optimization tools were performed to reduce the simulation time. (C) 2015 Elsevier B.V. All rights reserved.
Koleksiyonlar
- Makale [92796]