Verification and uniformity control of doses for Sr-90/ Y-90 intravascular brachytherapy sources using radiochromic film dosimetry
Tarih
2008Yazar
Ahmed, Asm Sabbir
DEMİR, Mustafa
Gurmen, Tevfi K.
DEMİR, Bayram
Babalik, Erhan
Üst veri
Tüm öğe kaydını gösterÖzet
Intravascular brachytherapy (IVBT) is a useful treatment modality for the recurrence of in-stent restenosis following drug-eluting stents (DES) or IVBT failure. The objective of this study was to measure the dose rate of Sr-90/Y-90 IVBT sources for comparison with that given by the manufacturer and to control the dose uniformities of these sources along the source axis. The dose rates of Sr-90/Y-90 beta sources were measured with a radiochromic film in a custom-made phantom. The films for calibration were irradiated using Co-60 photon beams. The results for the three sources were 4.5%, 2.3%, and 3.5% higher than the corresponding certificate values. Maximum and minimum of the dose rates varied within +/- 10% of those at source center; and maximum dose discrepancy for the first Sr-90/Y-90 source train was 8.2%; for the second source train, 7.1%; and for the third source train, 5.1%. Our study showed that the dose rates given by the manufacturer for the three Sr-90/Y-90 IVBT sources were reliable and dose uniformities were within +/- 10% along two thirds of the treatment length.