dc.date.issued2022
dc.identifier.citationŞENER H., ANILANMERT B., Mavis M. E. , Gursu G. G. , Cengiz S., "LC-MS/MS monitoring for explosives residues and OGSR with diverse ionization temperatures in soil & hands: 30 minutes for extraction plus elution", ACTA CHROMATOGRAPHICA, cilt.34, sa.3, ss.304-314, 2022
dc.identifier.issn1233-2356
dc.identifier.otherav_cf6bbfd5-4ab7-4e85-b689-53d181c09980
dc.identifier.othervv_1032021
dc.identifier.urihttp://hdl.handle.net/20.500.12627/184775
dc.identifier.urihttps://doi.org/10.1556/1326.2021.00930
dc.description.abstractA fast LC-APCI-MS/MS screening/confirmation method was developed and validated for trace analyses of 18 analytes which are explosives and organic gun shot residues including the challenging ones with diverse ionization conditions, in soil and on hands. (+) and (-) ionization modes were used after a single-step, low-volume solvent extraction procedure developed using methanol. Tape-lifting, stub, alcohol wipes, cotton bud were compared for collecting the residues from hands of a shooter. Tapelifting and stub gave the highest recoveries and tape-lifting was chosen. Gradient elution system using ammonium chloride:methanol was developed. Whole procedure lasted approximately 30 min, was validated in both matrices, applied to real samples as post-blast residues, smokeless powder and the hands of a shooter, after shooting. Most of the recoveries were >80% and since all the precisions were <15%, quantitation was possible for all. Limit of Detection (LOD) and Limit of Quantification (LOQ) values were: 0.2 54.1 and 0.3 190.0 ngg(-1) in soil, and 0.2 132.3 and 1.1 355.0 ngg(-1) in tape-lift.
dc.language.isoeng
dc.subjectGeneral Chemistry
dc.subjectPhysical Sciences
dc.subjectAnalytical Chemistry
dc.subjectChemistry (miscellaneous)
dc.subjectFiltration and Separation
dc.subjectTemel Bilimler
dc.subjectAnalitik Kimya
dc.subjectTemel Bilimler (SCI)
dc.subjectKimya
dc.subjectKİMYA, ANALİTİK
dc.titleLC-MS/MS monitoring for explosives residues and OGSR with diverse ionization temperatures in soil & hands: 30 minutes for extraction plus elution
dc.typeMakale
dc.relation.journalACTA CHROMATOGRAPHICA
dc.contributor.departmentKütahya Sağlık Bilimleri Üniversitesi , Mühendislik Ve Doğa Bilimleri Fakültesi , Adli Bilimler Bölümü
dc.identifier.volume34
dc.identifier.issue3
dc.identifier.startpage304
dc.identifier.endpage314
dc.contributor.firstauthorID3434088


Files in this item

FilesSizeFormatView

There are no files associated with this item.

This item appears in the following Collection(s)

Show simple item record