Basit öğe kaydını göster

dc.contributor.authorBILGE, Hatice Bilge
dc.contributor.authorKUCUCUK, Halil
dc.contributor.authorCAKIR, A
dc.contributor.authorFayda, EMRE MERDAN
dc.contributor.authorOKUTAN, Murat
dc.contributor.authorDADASBILGE, Alpar
dc.date.accessioned2021-03-04T07:37:45Z
dc.date.available2021-03-04T07:37:45Z
dc.date.issued2005
dc.identifier.citationCAKIR A., BILGE H. B. , DADASBILGE A., KUCUCUK H., OKUTAN M., Fayda E. M. , "The comparison of Co-60 and 4 MV photons matching dosimetry during half-beam technique", RADIATION MEASUREMENTS, cilt.39, sa.4, ss.447-450, 2005
dc.identifier.issn1350-4487
dc.identifier.othervv_1032021
dc.identifier.otherav_5f34e130-df6d-4261-a402-f5c8987d3b58
dc.identifier.urihttp://hdl.handle.net/20.500.12627/66523
dc.identifier.urihttps://doi.org/10.1016/j.radmeas.2004.10.001
dc.description.abstractIn this phantom study, we tried to compare matching dosimetry differences between half-blocking of Co-60 and asymmetric collimation of the 4 MV photons during craniospinal irradiation. The dose distributions are compared and discussed. Firstly, some gaps with different sizes are left between cranial and spinal field borders. Secondly, the fields are overlapped in the same sizes. We irradiate the films located in water-equivalent solid phantoms with Co-60 and 4MV photon beams. This study indicates that the field placement errors in +/- 1 mm are acceptable for both Co-60 and 4 MV photon energies during craniospinal irradiation with half-beam block technique. Within these limits the dose variations are specified in +/- 5%. However, the setup errors that are more than I mm are unacceptable for both asymmetric collimation of 4 MV photon and half-blocking of Co-60. (C) 2004 Elsevier Ltd. All rights reserved.
dc.language.isoeng
dc.subjectNükleer Fizik
dc.subjectTemel Bilimler (SCI)
dc.subjectNÜKLEAR BİLİMİ VE TEKNOLOJİSİ
dc.subjectFizik
dc.subjectTemel Bilimler
dc.titleThe comparison of Co-60 and 4 MV photons matching dosimetry during half-beam technique
dc.typeMakale
dc.relation.journalRADIATION MEASUREMENTS
dc.contributor.department, ,
dc.identifier.volume39
dc.identifier.issue4
dc.identifier.startpage447
dc.identifier.endpage450
dc.contributor.firstauthorID74659


Bu öğenin dosyaları:

DosyalarBoyutBiçimGöster

Bu öğe ile ilişkili dosya yok.

Bu öğe aşağıdaki koleksiyon(lar)da görünmektedir.

Basit öğe kaydını göster