Show simple item record

dc.contributor.authorYucel, Derya
dc.contributor.authorFayda, Merdan
dc.contributor.authorYilmaz, Sercan
dc.contributor.authorTatli, Hamza
dc.date.accessioned2021-03-04T14:22:20Z
dc.date.available2021-03-04T14:22:20Z
dc.identifier.citationTatli H., Yucel D., Yilmaz S., Fayda M., "Development of Independent MU/Treatment Time Verification Algorithm for Non-IMRT Treatment Planning: A Clinical Experience", 33rd International Physics Congress of the Turkish-Physical-Society (TPS), Bodrum, Türkiye, 6 - 10 Eylül 2017, cilt.1935
dc.identifier.othervv_1032021
dc.identifier.otherav_81353020-2d19-40ee-9411-c501eb7e476e
dc.identifier.urihttp://hdl.handle.net/20.500.12627/88081
dc.identifier.urihttps://doi.org/10.1063/1.5025987
dc.description.abstractThe aim of this study is to develop an algorithm for independent MU/treatment time (TT) verification for non-IMRT treatment plans, as a part of QA program to ensure treatment delivery accuracy. Two radiotherapy delivery units and their treatment planning systems (TPS) were commissioned in Liv Hospital Radiation Medicine Center, Tbilisi, Georgia. Beam data were collected according to vendors' collection guidelines, and AAPM reports recommendations, and processed by Microsoft Excel during in-house algorithm development. The algorithm is designed and optimized for calculating SSD and SAD treatment plans, based on AAPM TG114 dose calculation recommendations, coded and embedded in MS Excel spreadsheet, as a preliminary verification algorithm (VA). Treatment verification plans were created by TPSs based on IAEA TRS 430 recommendations, also calculated by VA, and point measurements were collected by solid water phantom, and compared. Study showed that, in-house VA can be used for non-IMRT plans MU/TT verifications.
dc.language.isoeng
dc.subjectDisiplinlerarası Fizik ve İlgili Bilim ve Teknoloji Alanları
dc.subjectTemel Bilimler
dc.subjectTemel Bilimler (SCI)
dc.subjectFizik
dc.subjectFİZİK, MULTİDİSİPLİNER
dc.titleDevelopment of Independent MU/Treatment Time Verification Algorithm for Non-IMRT Treatment Planning: A Clinical Experience
dc.typeBildiri
dc.contributor.departmentLiv Hosp Radiat Med Ctr , ,
dc.identifier.volume1935
dc.contributor.firstauthorID152772