Basit öğe kaydını göster

dc.contributor.authorInce, Z
dc.contributor.authorOrs, R
dc.contributor.authorOzek, E
dc.contributor.authorBekiroglu, N
dc.contributor.authorBilgen, H
dc.date.accessioned2021-03-04T18:43:26Z
dc.date.available2021-03-04T18:43:26Z
dc.date.issued1998
dc.identifier.citationBilgen H., Ince Z., Ozek E., Bekiroglu N., Ors R., "Transcutaneous measurement of hyperbilirubinaemia: comparison of the Minolta jaundice meter and the Ingram icterometer", ANNALS OF TROPICAL PAEDIATRICS, cilt.18, ss.325-328, 1998
dc.identifier.issn0272-4936
dc.identifier.otherav_8c376bc5-8bf4-4f70-b966-20f645e0c611
dc.identifier.othervv_1032021
dc.identifier.urihttp://hdl.handle.net/20.500.12627/94872
dc.identifier.urihttps://doi.org/10.1080/02724936.1998.11747968
dc.description.abstractThe effectiveness of two different non-invasive transcutaneous bilirubin measurement devices was compared with serum bilirubin levels in 96 healthy newborns. Transcutaneous measurements were obtained with the Minolta Air Shields jaundice meter and the Ingram icterometer and serum bilirubin levels were determined by a direct spectrophotometric method (Bilitron 444). A linear correlation existed between serum bilirubin values and the readings on both the Minolta jaundice meter (r= 0.83) and the Ingram icterometer (r = 0.78). The Kappa coefficient was 0.66. The sensitivity, specificity and positive and negative predictive values were 100%, 56%, 33% and 100% for the Minolta jaundice meter and 100%, 48%, 29% and 100% for the Ingram icterometer, respectively. The high sensitivity and negative predictive value of both devices render them suitable for screening neonatal hyperbilirubinaemia. However, because of its low cost, the Ingram icterometer is preferable to the more complex and expensive Minolta jaundice meter, especially in countries with a high birth rate, such as Turkey.
dc.language.isoeng
dc.subjectPharmaceutical Science
dc.subjectPediatrics
dc.subjectPediatrics, Perinatology and Child Health
dc.subjectLife Sciences
dc.subjectHealth Sciences
dc.subjectDahili Tıp Bilimleri
dc.subjectÇocuk Sağlığı ve Hastalıkları
dc.subjectSağlık Bilimleri
dc.subjectTıp
dc.subjectTROPİKAL TIP
dc.subjectKlinik Tıp (MED)
dc.subjectKlinik Tıp
dc.subjectPEDİATRİ
dc.titleTranscutaneous measurement of hyperbilirubinaemia: comparison of the Minolta jaundice meter and the Ingram icterometer
dc.typeMakale
dc.relation.journalANNALS OF TROPICAL PAEDIATRICS
dc.contributor.department, ,
dc.identifier.volume18
dc.identifier.issue4
dc.identifier.startpage325
dc.identifier.endpage328
dc.contributor.firstauthorID2368912


Bu öğenin dosyaları:

DosyalarBoyutBiçimGöster

Bu öğe ile ilişkili dosya yok.

Bu öğe aşağıdaki koleksiyon(lar)da görünmektedir.

Basit öğe kaydını göster